Surface Finish Requirements for Medical Implant Bores

Medical implant bores need smooth surface finishes for biocompatibility and fatigue life. A rough bore surface causes stress concentrations that lead to implant failure under cyclic loading. I’ve worked on hip stems, trauma nails, bone screws, and surgical instruments over the years, and the surface finish requirements vary by application. The general range I see is Ra 0.4-0.8um for bone screws and trauma nails and Ra 0.2-0.4um for load-bearing implantable devices. Surgical instruments are less demanding at Ra 0.8um. But these numbers don’t tell the full story — the finish needs to be consistent across the full bore length, not just at the entry. ...

June 12, 2026 · Engineer at the Deep Hole

Deep Hole Drilling for Medical Implants: Small Diameters, Tight Tolerance

Medical implant drilling is a specialized corner of deep hole drilling. The diameters are small — 1mm to 6mm — and the depth-to-diameter ratios can go to 400:1. The materials are titanium and surgical stainless steel. And the tolerance requirements are tighter than anything else I’ve worked with. The parts I’ve worked on are bone screws, trauma nails, and surgical instruments. They all need a straight, clean hole through the full length. A bone screw that’s 4mm in diameter and 120mm long needs a hole through the center that’s maybe 1.5mm. The wall thickness around that hole is about 1.25mm. There’s no room for error. ...

June 11, 2026 · Engineer at the Deep Hole